Polyethylene Terephthalate Non-Intentionally Added Substance Migration Testing Protocols

Polyethylene terephthalate NIAS verification requires multi-platform chromatographic screening down to 0.01 mg/kg backed by toxicological threshold alignment.

24.10.26 18 min

Matrix

Polyethylene terephthalate resin placed in food contact yields chemical migrants that never appeared on an additive specification. Commercial polyester containers release cyclic oligomers, thermal breakdown fragments, and catalytic residue under standard migration conditions. Laboratory screening frequently encounters substances lacking specific migration limits under Commission Regulation (EU) 10/2011.

Testing for non-intentionally added substances begins by mapping the base polymer condensation reaction alongside thermal history. Ethylene glycol and purified terephthalic acid polymerize under vacuum at temperatures exceeding 270 degrees Celsius. This environment drives side reactions that generate diethylene glycol, acetaldehyde, and low molecular weight cyclic oligomers.

Analytical verification requires direct quantification of the primary cyclic trimer, known chemically as cyclic tris(ethylene terephthalate). This trimer constitutes the dominant migrant from virgin polyethylene terephthalate, often representing over eighty percent of total oligomeric migration. The substance possesses a molecular weight of 576.5 grams per mole, placing it well below the 1,000 Dalton absorption threshold for gastrointestinal uptake.

Testing protocols rely on food simulants specified in European Union regulations: ten percent ethanol for aqueous foods, three percent acetic acid for acidic media, and vegetable oil or poly(2,6-diphenyl-p-phenylene oxide), known commercially as Tenax, for dry storage. Acetic acid at three percent creates an aggressive hydrolytic environment at elevated temperatures, accelerating ester cleavage.

Ten percent ethanol held for ten days at sixty degrees Celsius yields measurable cyclic trimer concentrations that increase linearly with contact surface area.

Migration testing must evaluate the finished article rather than the resin pellet. Extrusion and stretch blow molding introduce shear and thermal stresses that alter the migrant profile. Preform manufacture subjects the polyester to barrel temperatures between 280 and 300 degrees Celsius.

These conditions cleave ester linkages, generating volatile aldehydes and vinyl esters. Residual moisture accelerates hydrolytic degradation, producing additional linear and cyclic oligomeric fragments. Analytical protocols that examine only virgin granulate miss volatile breakdown products created during stretch blow molding.

Testing laboratories operate with defined limits of quantification. A non-detectable result means only that the instrument did not register a signal above its baseline noise at the target mass-to-charge ratio. High-resolution mass spectrometry screening carries a screening detection limit, typically fixed at 0.01 milligrams per kilogram of food simulant.

Any migrant appearing below this value escapes mandatory structural identification when toxicological evaluation demonstrates absence of genotoxicity. Analytical chemists document the exact calibration curve, ionization mode, and mobile phase parameters supporting this boundary.

Testing protocols follow a strict sequence to capture volatile, semi-volatile, and non-volatile fractions across processing stages:

  1. Raw pellet degassing measures initial volatile headspace concentrations before processing, establishing a baseline for acetaldehyde and ethylene glycol levels.
  2. Melt process logging records thermal residence times across extruder zones, identifying points where excessive heat accelerates transesterification side reactions.
  3. Finished container exposure subjects fabricated bottles or sheets to target food simulants inside certified migration cells, preventing edge leakage and non-contact surface extraction.
  4. Extract concentration evaporation reduces simulant volumes under gentle nitrogen flow without exceeding forty degrees Celsius, preserving volatile breakdown fractions for analytical injection.

A testing laboratory selects contact conditions reflecting the intended shelf life. Storage exceeding six months at room temperature demands accelerated testing at sixty degrees Celsius for ten days. Hot-fill operations require testing at one hundred degrees Celsius for two hours, followed by ambient storage duration.

When testing conditions fail to reflect the filling process, calculated exposure models fail regulatory scrutiny.

A supplier often claims that virgin resin certificates cover the compliance of finished thermoformed trays without additional migration data.

Melt

Thermal processing converts stable polymer chains into active degradation sites. Polyethylene terephthalate degrades through random chain scission, thermal oxidation, and transesterification. Extruders generate local shear heating, raising melt temperatures twenty degrees above setpoint values.

Moisture content above fifty parts per million drives immediate hydrolytic cleavage of ester bonds. This reaction produces terephthalic acid, monohydroxyethyl terephthalate, and shorter chain fragments. These low molecular weight fragments migrate rapidly into fatty and alcoholic foods.

Catalyst systems left within the resin drive further chemical alterations during melting. Antimony trioxide remains the standard condensation catalyst, typically present between 150 and 300 milligrams per kilogram of resin. Titanium and germanium formulations serve as commercial alternatives.

Residual antimony atoms coordinate with carbonyl oxygens in the polyester backbone, catalyzing cyclic oligomer formation during bottle blowing. Migration testing protocols must account for both antimony migration, governed by a specific migration limit of 0.04 milligrams per kilogram under Regulation (EU) 10/2011, and the organic species generated by catalyst-assisted degradation.

Thermal Degradation Byproducts and Migration Behavior in Polyethylene Terephthalate Processing
Substance Identity Chemical Structure Class Primary Formation Mechanism Volatilization Temperature (C) Common Detection Simulant
Acetaldehyde Aliphatic aldehyde Thermal cleavage of vinyl ester ends 20.2 Headspace gas chromatography
Cyclic trimer Cyclic polyester ester Intramolecular transesterification 320 to 325 Ethanol 95 percent and isooctane
2-Methyl-1,3-dioxolane Cyclic acetal Acetaldehyde and ethylene glycol reaction 82 to 84 Ten percent ethanol
Terephthalic acid monoethylene glycol ester Aromatic ester acid Hydrolytic scission of polymer chain Decomposes above 200 Three percent acetic acid
Vinyl terephthalate Vinyl ester Beta-elimination during extrusion 145 to 150 Poly(2,6-diphenyl-p-phenylene oxide)

Acetaldehyde generation represents a visible degradation pathway in polyester conversion. It forms when hydroxyethyl end groups undergo scission, creating a vinyl ester intermediate and an acetaldehyde molecule. Acetaldehyde imparts an off-flavor to mineral water at concentrations as low as ten to twenty micrograms per liter.

The European Union lists acetaldehyde as an authorized monomer under Substance Table 1 of Regulation (EU) 10/2011, assigning a specific migration limit of six milligrams per kilogram. Converters frequently introduce acetaldehyde scavengers, such as 2-aminobenzamide, to control organoleptic defects. Scavengers introduce distinct reaction byproducts that classify as non-intentionally added substances.

The reaction between 2-aminobenzamide and acetaldehyde generates 1,2,3,4-tetrahydro-2-methyl-4-quinazolinone. This cyclic adduct migrates into water and aqueous simulants. Testing laboratories must synthesize analytical standards to quantify this compound, since commercial databases lack validated calibration responses for proprietary scavenger complexes.

Unreacted anthranilamide and its oxidation products also enter the migration extract, demanding separate chromatographic separation and toxicological evaluation.

Addition of two hundred parts per million of aminobenzamide reduces acetaldehyde headspace levels while generating anthranilamide reaction complexes that migrate into three percent acetic acid.

Solid-state polymerization alters the migrant profile by driving off volatile fragments under high temperature and vacuum. Resin intended for bottle blowing undergoes solid-state treatment to raise intrinsic viscosity from 0.60 to 0.84 deciliters per gram. This thermal conditioning reduces volatile monomer concentrations while increasing crystalline content.

Mechanical properties improve, yet high intrinsic viscosity materials demand higher processing temperatures during subsequent stretch blow molding. Higher melt temperatures regenerate cyclic oligomers, demonstrating that upstream thermal stripping cannot prevent downstream chemical generation.

A clear polymer film loop extends between two sensor jaws mounted on black metal frames inside an industrial production facility.

How Do Degradation Products Behave in Acidic Simulants?

Acidic food simulants accelerate ester hydrolysis at the container interface. Three percent acetic acid maintained at seventy degrees Celsius attacks terminal ester linkages, shifting the oligomeric equilibrium. Linear oligomers bearing terminal carboxyl groups increase in concentration relative to cyclic species.

Hydrolysis rate constants depend directly on storage temperature, following Arrhenius behavior with activation energies between eighty and one hundred kilojoules per mole. Laboratory reports must state whether migration testing used overall migration simulant exposure or single-sided contact cells, because submerged testing extracts oligomers from exterior trimmed edges that never touch food during commercial distribution.

Processing scrap recycling generates a compounding loop of degradation products. Regrind added back into the feed throat carries degraded functional groups, residual colorants, and degraded scavengers. Each extrusion cycle shears the polymer chain, increasing the concentration of vinyl end groups and carboxylic acid termini.

The acid termini act as autocatalytic sites for further hydrolytic cleavage during subsequent melting. Finished articles containing thirty percent industrial regrind show double the cyclic dimer and linear trimer migration levels compared to articles produced entirely from virgin granulate.

A converter substituting alternate resin grades without re-running migration verification risks introducing unquantified scavenger adducts into commercial distribution chains.

Spectrum

Screening for non-intentionally added substances requires multi-detector chromatographic platforms. Single-instrument protocols fail because volatile solvents, process stabilizers, cyclic oligomers, and mineral oils span broad boiling point and polarity ranges. Gas chromatography coupled with mass spectrometry targets volatile and semi-volatile migrants with molecular weights below five hundred Daltons.

Liquid chromatography coupled to high-resolution mass spectrometry captures non-volatile, polar, and high molecular weight oligomers. Testing facilities run both platforms on a single simulant extract to achieve complete characterization.

Gas chromatography utilizes splitless injection or programmable temperature vaporization to introduce simulant concentrates onto non-polar capillary columns. Headspace sampling methods evaluate volatile aldehydes without solvent extraction interference. For semi-volatile species, extract concentration under nitrogen precedes injection.

Mass spectrometers operate in electron ionization mode at seventy electron volts, generating fragmentation spectra that match against commercial mass spectral libraries. Chromatographers verify spectral matches by comparing experimental retention indices against homologous n-alkane series. Matches below eighty-five percent spectral similarity require manual deconvolution and structural confirmation through high-resolution accurate mass measurement.

Analytical Screening Configurations for Comprehensive Food Contact Migration Profiling
Instrument Configuration Target Compound Class Ionization / Detection Mode Typical Column Stationary Phase Linear Dynamic Range
Headspace GC-MS Volatile aldehydes and solvents Electron ionization (70 eV) 5 percent phenyl polysilphenylene-siloxane 0.005 to 5.0 mg/kg
Liquid Injection GC-MS/MS Plasticizers, slip agents, scavengers Positive chemical ionization 5 percent diphenyl 95 percent dimethylpolysiloxane 0.010 to 10.0 mg/kg
UHPLC-ESI-Q-TOF-MS Cyclic oligomers, photoinitiators Electrospray positive and negative Octadecyl carbon-18 (sub-2-micron) 0.001 to 2.5 mg/kg
UHPLC-CAD Non-chromophoric linear oligomers Charged aerosol detection Hydrophilic interaction liquid chromatography 0.050 to 20.0 mg/kg

Liquid chromatography utilizes ultra-high-performance systems connected to quadrupole time-of-flight or orbitrap mass analyzers. Electrospray ionization operates across both positive and negative polarity switches to capture acidic and basic functional groups. Accurate mass errors below five parts per million allow unequivocal determination of elemental formulas.

Chromatographic separation of cyclic polyester oligomers resolves series from the cyclic dimer up to the cyclic octamer. The cyclic trimer, cyclic tetramer, and cyclic pentamer appear as distinct chromatographic peaks, accompanied by ammonium or sodium adduct ions depending on mobile phase additive composition.

Quantification of unknown peaks presents severe analytical hurdles when certified reference materials do not exist. Laboratories calibrate their detector response against representative proxy standards. For gas chromatography, internal standards like deuterated benzophenone or diethyl phthalate provide semi-quantitative concentration calculations.

In liquid chromatography, di-tert-butyl-4-hydroxytoluene or a purified polyester cyclic trimer standard establishes response factors. The assumption of uniform ionization efficiency introduces quantification uncertainty. Electrospray response factors vary by up to two orders of magnitude between different molecular structures, meaning an uncalibrated concentration estimate can understate actual migration by an order of magnitude.

An insulated stainless steel tumbler is firmly held by a custom-padded vise within a clear industrial testing enclosure.

Will Gas Chromatography Detect Nonvolatile Oligomeric Cyclic Esters?

Standard gas chromatography fails to resolve polyester oligomers above the cyclic trimer. Thermal desorption or inlet temperatures above 320 degrees Celsius induce pyrolysis inside the injector liner. Pyrolysis splits high molecular weight oligomers into lower mass fragments, creating false peaks that misrepresent actual migrant distribution.

High-temperature gas chromatography using aluminum-clad or metal capillary columns extends the operational window up to four hundred degrees Celsius, enabling elution of the cyclic trimer and tetramer. For the cyclic pentamer through octamer, liquid chromatography remains the sole reliable quantitative method.

Non-intentionally added substance identification follows a tiered evaluation procedure:

  • Initial chromatographic deconvolution isolates overlapping peaks from baseline noise and column bleed components using deconvolution algorithms.
  • Blank subtraction screening eliminates background contaminants originating from laboratory glassware, extraction solvents, and solid-phase extraction cartridges.
  • Accurate mass elemental assignment establishes chemical formulas for all peaks exceeding the 0.01 milligram per kilogram analytical threshold using isotopic abundance patterns.
  • Structural database cross-referencing queries spectral libraries, structural databases, and polymer additive registers to propose candidate molecular architectures.

Accurate quantification requires rigorous matrix effect corrections. Food simulants like Tenax and vegetable oil retain hydrophobic migrants, reducing extraction yields during solvent back-extraction. Testing laboratories determine analyte recovery rates through standard addition protocols, adding known quantities of surrogates prior to incubation.

Recovery rates below seventy percent invalidate the test report under ISO/IEC 17025 quality frameworks.

Analytical quantification without matrix-matched calibration standards produces systemic uncertainty that authority inspectors reject during formal border import verifications.

Screening protocols must account for blind spots inherent to electrospray ionization. Highly non-polar hydrocarbons, including saturated oligomeric fragments and low molecular weight polyolefins from blending agents, exhibit poor ionization efficiency in electrospray fields. Atmospheric pressure chemical ionization bridges this gap for medium polarity migrants.

Failure to pair electrospray with atmospheric pressure chemical ionization or gas chromatography permits entire classes of non-polar lubricants and defoamers to pass undetected through laboratory validation screens.

Omitting atmospheric pressure chemical ionization from analytical verification cascades leaves undetected non-polar lubricants in the migration extract.

A clear polymer container assembly connects to a metallic test fixture positioned beneath an industrial press within a dark workshop.

Threshold

Evaluating unidentified migrants requires established toxicological assessment trees. When a chromatographic peak cannot be matched to a reference standard with confirmed toxicological data, safety assessment relies on the Threshold of Toxicological Concern concept. The European Food Safety Authority and the United States Food and Drug Administration accept this approach to benchmark human exposure.

Testing laboratories assign substances into Cramer Classes using the revised decision tree algorithm based on chemical structure, functional groups, and metabolic fate models.

Cramer Class I covers simple chemical structures with known metabolic pathways and low oral toxicity. The human exposure threshold for Class I is 1,800 micrograms per person per day. Class II includes intermediate structures that show less rapid metabolism, carrying an exposure limit of 540 micrograms per person per day.

Class III covers complex chemical arrangements, reactive functional groups, or structures containing elements other than carbon, hydrogen, oxygen, and nitrogen, with a daily exposure limit of 90 micrograms per person per day. For substances carrying structural alerts for genotoxicity, the Threshold of Toxicological Concern drops to 0.15 micrograms per person per day, equivalent to a dietary concentration of 0.0025 micrograms per kilogram of food.

Converting daily exposure thresholds into analytical migration limits requires standard packaging assumptions. European regulations assume that a person weighing sixty kilograms consumes one kilogram of food packaged in six square decimeters of surface area daily. Under this assumption, a Cramer Class III threshold of ninety micrograms per person per day translates directly to an analytical migration limit of 0.090 milligrams per kilogram of food or food simulant.

For infant nutrition, food consumption ratios per kilogram of body weight are higher, depressing acceptable migration limits by a factor of three to five.

Toxicological Thresholds and Calculated Migration Allocation Limits by Structural Class
Toxicological Classification Exposure Threshold (ug/person/day) Standard Food Limit (mg/kg food) Infant Food Limit (mg/kg food) Required Analytical LOQ (mg/kg)
Cramer Class I (Low toxicity) 1800 1.800 0.360 0.050
Cramer Class II (Intermediate) 540 0.540 0.108 0.020
Cramer Class III (High toxicity) 90 0.090 0.018 0.010
Organophosphates / Carbamates 18 0.018 0.003 0.002
Genotoxicity Alert Positive 0.15 0.00015 0.00003 0.0001

When high-resolution mass spectrometry detects an unknown peak above 0.01 milligrams per kilogram, the analytical laboratory must screen the structure for mutagenic structural alerts. In silico quantitative structure-activity relationship models, such as Derek Nexus or the OECD QSAR Toolbox, evaluate potential reactivity toward genetic material. If the software flags an aromatic amine, epoxide, alkyl halide, or alpha-beta-unsaturated carbonyl group, the chemist cannot apply standard Cramer Class thresholds.

The migrant must be proven non-genotoxic through experimental bioassays or held below the 0.00015 milligram per kilogram threshold.

In vitro bioassays provide direct empirical biological data for complex migration extracts containing multiple unidentified compounds. The bacterial reverse mutation assay, known as the Ames test, screens for gene mutations across Salmonella typhimurium and Escherichia coli strains. Standard testing exposes test strains to concentrated migration extracts up to non-cytotoxic limits.

Extracts derived from recycled polyester require concentration by solid-phase extraction to bring trace migrants into the dynamic response range of the assay, introducing extraction recovery variables that must be tracked through surrogate standards.

Articles yielding migration extracts that trigger structural alerts for bacterial mutagenicity fail compliance evaluations regardless of overall migration weight metrics.

Calculating consumer exposure requires validated contact surface area data. A 500-milliliter single-serve bottle maintains a surface area of approximately four square decimeters, yielding an area-to-volume ratio of eight square decimeters per liter. A five-gallon water cooler bottle maintains an area-to-volume ratio of two square decimeters per liter.

Sourcing teams cannot transpose laboratory reports generated on large containers to small single-serve formats without recalculating the volumetric migrant distribution. Applying large container compliance certificates to small portion packaging constitutes an auditing error.

Post-consumer recycled polyethylene terephthalate alters the risk profile. Mechanical recycling introduces cleaning chemical residue, flavor compounds, and thermal degradation markers from prior consumer lifecycles. Common contaminants include limonene, p-cymene, and benzophenone from printed labels.

Decontamination processes operating under European Regulation (EU) 2022/1616 require challenge testing with surrogate contaminants to prove cleaning efficiencies exceeding ninety-nine percent. Sourcing managers must obtain documentation certifying that decontamination technology cleans volatile and semi-volatile migrants below the Threshold of Toxicological Concern.

Can a buyer accept declarations citing structural similarity to clear Class III compounds without direct bioassay screening? Structural analogues provide supporting evidence only when functional groups and metabolic routes align across metabolic models. Regulators reject toxicological justifications based on loose generic comparisons when sterically hindered groups or active metabolites alter biological endpoints.

A metallic pan holds a pale, viscous polymer material on a processing unit within a cleanroom environment.

Recourse

Conformity documentation must connect individual raw material batches to finished article test reports. A Declaration of Conformity issued under Regulation (EC) 1935/2004 and Regulation (EU) 10/2011 is not a product brochure. It is a legal instrument that transfers liability across the supply chain.

Sourcing managers must demand comprehensive supporting dossiers that substantiate every statement on the declaration. When customs officials or food safety inspectors examine a shipment at a port of entry, an unsupported declaration leads to border rejection and seizure.

Auditing the Declaration of Conformity requires systematic verification of supporting annexes. Importers verify that the testing laboratories maintain ISO/IEC 17025 accreditation for specific migration test methods rather than general physical testing. The test report must specify the trade name, resin grade, and conversion process conditions matching the production run.

If the test report lists only virgin resin granules while the container contains twenty-five percent post-consumer material, the report has zero legal standing for the shipped lot.

Contractual agreements must allocate costs arising from non-compliant non-intentionally added substance migration. Sourcing teams insert technical compliance schedules into purchasing agreements to protect capital. The contract specifies that suppliers warrant finished articles against unauthorized migrants above defined toxicological thresholds.

If laboratory re-testing reveals cyclic oligomers or degradation fragments exceeding regulatory limits, the supplier assumes the landed cost, warehousing fees, and disposal expenses.

A rigorous compliance auditing checklist screens documentation before cargo dispatch:

  • Batch traceability matching verifies that heat numbers and lot codes on shipping documents correspond to the analytical test report references.
  • Simulant exposure alignment confirms that testing protocols utilized the exact simulants and temperature-time combinations mandated for the food filling conditions.
  • Substance authorization verification cross-checks every declared additive against Union lists, confirming compliance with individual specific migration limits.
  • Screening detection limit audit ensures that non-intentionally added substance evaluations applied an analytical screening detection limit of 0.01 milligrams per kilogram.
  • Recycled resin authorization inspects recycling plant decontamination approval numbers registered under European Union registers for recycled plastic production.

Consider a sourcing scenario involving 100,000 thermoformed polyethylene terephthalate containers intended for ambient retail deli packaging. Sourcing teams specify testing with simulant D1 (fifty percent ethanol) for ten days at forty degrees Celsius to cover dairy salads. A supplier submits a legacy report citing simulant A (ten percent ethanol) tested for ten days at forty degrees Celsius.

Accepting the incorrect simulant invalidates compliance. Dairy matrices contain lipophilic fats that swell the polyester matrix, accelerating oligomer diffusion rates up to five times faster than aqueous simulants. Importing containers backed by an incompatible simulant report leaves the brand owner exposed to mandatory product recalls and regulatory fines.

Testing fees and toxicological assessments represent substantial landed costs. High-resolution mass spectrometry screening paired with gas chromatography and liquid chromatography ranges between 2,500 and 4,500 euros per sample. If unknown peaks trigger Cramer Class III evaluations or in vitro bioassays, testing costs escalate by an additional 3,000 to 6,000 euros.

Sourcing organizations factor these analytical fees into their annual supplier qualification budgets. Omitting screening protocols saves initial cash flow while exposing the business to border rejections, inventory write-offs, and product liability litigation.

Commercial contracts define clear warranty provisions for packaging compliance. A standard clause states: “The supplier warrants that the supplied articles comply with Regulation (EC) 1935/2004, Regulation (EC) 2023/2006, and Regulation (EU) 10/2011, and that non-intentionally added substances have been assessed in accordance with internationally recognized scientific principles. The supplier agrees to indemnify the buyer against all direct costs, recalls, testing expenses, and administrative penalties arising from non-compliant migration levels discovered in certified lots.”

Nomenclature

Structural Alerts

Meaning ~ Molecular substructures identified within a chemical entity correlate with specific toxicological or reactivity outcomes during polymer processing.

Post-Consumer Recycled PET

Meaning ~ Thermoplastic polyester derived from waste packaging streams functions as a regulated raw material for moulding operations that demand specific thermal and mechanical performance thresholds.

Regulation EC 1935 2004

Meaning ~ Framework European legislation mandates that materials contacting food must not transfer constituents in quantities that endanger human health or unacceptably alter food composition.

Mass Spectrometry Screening

Meaning ~ General laboratory methods used to survey the breadth of volatile and non volatile substances migrating from polymeric surfaces identify non intentionally added substances without specific pre selection.

High-Resolution Mass Spectrometry

Meaning ~ Analytical instruments that measure the mass-to-charge ratio of ions with high precision allow for the identification of unknown chemical compounds in complex mixtures.

Cramer Classification

Meaning ~ This protocol defines the standard for assessing polymer thermal degradation during injection moulding cycles.

Regulation EU 2022 1616

Meaning ~ Legislative frameworks establish the safety and traceability requirements for recycled plastic materials intended for food contact within the European market.

Solid State Polymerization

Meaning ~ Solid state polymerization is a thermal upgrading process applied below the melting point of a resin to increase molecular weight and remove residual volatiles from preformed pellets.

Specific Migration Limit

Meaning ~ Quantitative thresholds define the maximum permitted amount of a particular substance that can transfer from a finished plastic part into a food product or simulant.

Threshold of Toxicological Concern

Meaning ~ A quantitative exposure exposure limit identifies the maximum quantity of a chemical migration into a food contact polymer that avoids chronic health risks regardless of the specific chemical structure.

Migration Limit

Meaning ~ A maximum concentration threshold represents the legal limit for specific chemical substances migrating from food contact polymers into the contained foodstuff.

Toxicological Concern

Meaning ~ Risk assessment framework establishes the probability of adverse human health outcomes provoked by human exposure to chemical substances identified in consumer goods.