Meaning
Microscopic pockets of gas adhere to the surface of a solid specimen when it is submerged in a testing liquid. During density measurements of moulded polymer parts, micro-bubbles introduce a notable source of buoyancy error that artificially reduces the calculated density. These tiny bubbles can go unnoticed by the naked eye but can cause a high-quality polymer to appear out of specification.
Their presence is often driven by surface roughness or the high surface tension of the liquid, and their volume alters the apparent mass of the sample in the liquid.
Buoyancy Disturbance
The physical displacement of liquid by the trapped gas alters the weight of the sample in the fluid. Even a few micro-bubbles can displace enough liquid to mimic a void in the material, leading to incorrect reject decisions.
Surface Adhesion
Chemical interactions between the plastic surface and the liquid influence the formation of these gas pockets. Hydrophobic polymers like polypropylene are particularly prone to attracting micro-bubbles when submerged in water. This attraction occurs because the water does not wet the polymer surface easily, leaving air trapped in the microscopic surface crevices.
Eradication Method
Preventive steps can be taken to eliminate these gas pockets before taking a measurement. Adding a surfactant to the water or pre-wetting the polymer sample with an alcohol solution can prevent micro-bubbles from forming. Ultrasonic cleaning of the submerged specimen also forces any trapped air to detach from the surface.