Meaning
Calorimetric measurement accuracy depends on uniform heat transfer across the physical interface between the specimen container and differential scanning calorimetry sensor platform. Optimal sample pan thermal contact minimizes thermal resistance, preventing artifact peak broadening and artificial temperature delays during dynamic testing runs. Flat-bottomed aluminum pans ensure high contact surface area against the thermoelectric sensor face.
Curved or crimped pan bottoms distort measured heat flow signals.
Interface Resistance
Microscopic surface roughness or trapped debris between the pan bottom and sensor plate restricts thermal conductive flow. Poor sample pan thermal contact causes apparent thermal lag, shifting measured glass transition temperatures higher than actual material values. Clean sensor surfaces ensure consistent thermal conductivity.
Pan Geometry
Deformed or uneven sample pan bottoms reduce effective contact area, restricting heat flow rates into polymer specimens. Insufficient sample pan thermal contact broadens melting endotherms and distorts calculated crystallinity percentages. Using dedicated crimping presses ensures reproducible pan bottom flatness across testing samples.
Calibration Offset
Thermal calibration routines rely on sharp melting onset transitions of high-purity indium standard metals to establish temperature scale accuracy. Altering sample pan thermal contact between calibration runs and polymer testing introduces systematic temperature measurement errors. Maintaining uniform pan deformation practices ensures analytical accuracy across routine quality audits.