Meaning
Electrochemical potential differences that arise across the boundary between the reference electrode electrolyte and the sample solvent can distort acidity measurements of polymer solutions. This boundary voltage, known as liquid junction potential, fluctuates when the ionic strength of the sample varies during titration. In the testing of recycled plastics, these fluctuations obscure the exact titration endpoint of acidic degradation products.
Unreliable measurements allow resins with high acidic content to enter the production line, leading to mould corrosion.
Junction Behavior
Reference electrodes require a continuous flow of internal electrolyte through a porous junction to maintain electrical contact with the sample. When the sample consists of a non-aqueous polymer solution, the interface between the aqueous electrolyte and the organic solvent creates a large potential difference. This potential varies as the titration proceeds, introducing a systematic error into the measured pH value.
Using matching non-aqueous electrolytes minimizes this junction potential.
Electrode Selection
Selecting the appropriate junction style is essential for achieving reliable results in viscous polymer solutions. Standard ceramic junctions clog easily when polymer chains precipitate on the porous barrier. A clogged junction increases the resistance and causes the potential to drift uncontrollably.
Choosing a sleeve junction provides a steady flow of electrolyte that sweeps away the polymer, maintaining a stable potential.
Control Strategy
Stable reference potentials are achieved by using double-junction electrodes filled with ethanolic lithium chloride. This configuration isolates the reference element from the solvent.