Meaning
Reduction in the ionization efficiency of target analytes occurs when co-eluting compounds compete for charge during liquid chromatography mass spectrometry. This electrospray ionization matrix suppression alters the quantified signal of polymer additives like antioxidants and slip agents. Matrix components, such as low molecular weight oligomers, dominate the droplet surface and prevent analyte evaporation.
The phenomenon is restricted to electrospray methods and does not affect gas chromatography.
Additive Analysis
Extraction of additives from moulded plastic parts often brings along soluble polymer fragments. During electrospray ionization matrix suppression, these substances prevent the mass spectrometer from detecting the true additive concentration. Moulders rely on accurate additive assays to verify that incoming resins match specifications.
Ionization Efficiency
Competition for available protons or charge carriers on the microdroplet surface is the primary physical cause of signal loss. In the presence of electrospray ionization matrix suppression, the target molecules remain in the uncharged liquid state and fail to enter the mass analyzer. Analysts use deuterated internal standards or run sample dilutions to mitigate this ionization blockade.
Dilution reduces the absolute concentration of matrix interferences, which restores the signal of the additive of interest. This technique allows for the precise determination of stabilizer concentrations in complex polymer extracts.
Quality Control
Erroneous additive readings can lead to resin batches being wrongly rejected or accepted for production. When electrospray ionization matrix suppression goes undetected, the laboratory might report a deficit of stabilizers in a polymer lot. This incorrect report can prompt the addition of extra stabilizer, which elevates raw material costs and can cause blooming on the moulded part surface.
Robust chromatographic separation prevents these analytical errors by resolving the matrix from the analytes.